Note:
Ensure that the
       	    calibration solution
       	    covers the entire
       	    sensor cell and the 
       	    logger is agitated to
       	    release entrapped 
       	    air bubbles.

Retrieve the settings from the attached LTC Levelogger by clicking  .
. 
Stop the LTC Levelogger  by clicking  .
. 
Start the Solinst Levelogger Software Calibration Wizard by clicking the Conductivity Cal tab. The wizard will guide you through the calibration process. You can exit the wizard after any step to cancel the calibration session.
Introduction: The Solinst LTC Levelogger must be connected to the computer with an Optical Reader (Desktop Reader 5 or Field Reader 5) during the calibration and must be in the stopped mode. Ensure you use fresh calibration solution and allow thermal equilibration of the solution during the process. Also ensure no air bubbles are present on the sensor and that the pins are clean. See Section 10.2 for recommended methods to clean the sensor pins. See Section 10.1.5.3 for protection against biofouling conditions.

Figure 6-2 Levelogger Conductivity Calibration Wizard Introduction